Calibrating the software to your patient display

Monitor calibration

Letter Size Calibration

Its essential that your calibration is set correctly to ensure consistent VA measurements. During the initial setup as you go through the setup screens, you will need to input the height and the width of a plus sign. This is input in millimetres. It is imperative to measure these heights very accurately, and they need to be measured on the “PATIENT VIEWING SCREEN”, and NOT on your laptop, unless you are using your laptop as the patient viewing screen. This measurement tells Spectrum about your patient viewing monitor, as well  as the monitors resolution.

As you continue through the setup screens, please input “6000” as your patient viewing distance. This might not be the patient viewing distance, but we need to check that the letter sizes are correct, and this is the standard we use.

Scroll your mouse wheel till the 6/60 line on the “Logmar Chart” is displayed. You might need to Randomize the letters till you get a letter with a straight vertical limb Now go to your patient viewing monitor and measure the height of the letter on the screen, and the limb height should be 87.5 mm high. If this is the height of the letter, then your setup is correct. If the size is not correct, recheck the height and width in millimetres that you measured on the plus sign. You now need to go to the “Options Screen” on the top menubar, and select “Setup Wizard” and when you get to the screen where you are required to input the “room length”, then input your correct room length that you measure from your pheropter or trial frame when placed before a patients face, to your Patients viewing monitor. In order to guarantee accurate and consistent VA measurements, it is essential to carry out this exercise correctly.

In summary:

  1. Select Options from the top menubar
  2. Select setup wizard
  3. Go To “Step 2 of 4”
  4. Enter 6000 as your room length, and check the tickbox “using a mirror” if you are using one, leave unchecked if not.
  5. On “Step 3” select the Patients screen from the number displayed on the screen click “next”
  6. Measure in Millimetres the height and width of the large plus sign, (ON THE PATIENT MONITOR)  and enter the values
  7. Click “Finish”
  8. Select a 6/60 letter on the VA screen, (Patient Monitor) and measure the height of the letter
  9. Should be 87.5 mm if not, redo height and width of plus sign
  10. If it is 87.5, go back to step 2 and enter Room Length from the patient to the Monitor, and if using a Mirror, ADD this to the length from the mirror to the patient monitor., and check the Mirror tick box on the room length screen
  11. Select “finish” your letter size is correctly calibrated

Congratulations, Your Letter size is now correctly calibrated

Calibrate When Using a Mirror

Go through the normal Letter Calibration process, as above and when you have the limb height of the 6/60 letter correct on the monitor with the room length set as 6000, then go back to the setup wizard where you are required to enter the actual room length, from your pheropter or trial frame to the mirror, then add the distance from the mirror to the monitor above the patients head check the tick box that says “are you using a mirror” Its essential that your calibration is set correctly to ensure consistent VA measurements.

Congratulations, Your Letter size is now correctly calibrated and setup for use with a mirror.

New Practice re-calibrate

When moving to a new practice or test room, or if you are screening at a school, or if you have changed your “patient viewing monitor”, you will need to check and Letter Calibration process. as above

When screening at a school it will only be necessary to adjust your room length to the length you are using at the school.

As a refresher, here is the process:- Go to the “Options Screen” on the top menubar, and select “Setup Wizard” and when you get to the screen where you are required to input the “room length”, then input the distance you are using at the school from the pupil to the screen. In order to guarantee accurate and consistent VA measurements, it is essential to carry put this exercise correctly.

When returning to your practice, it is recommended that you run through the Letter Calibration process as above to just be sure of your measurements. If you are using a different monitor at the school, it will be necessary to go through the entire setup process on the monitor the pupils are viewing, as this might be different resolution to the monitor in your test room, or if they are viewing the laptop monitor directly it will be necessary to once again go through the Letter Calibration process as above, and then you will need to go through the process again on your return to your test room

Troubleshooting

Screens swopped

This is a common problem when installing other programs on your computer, and then the basic settings are disrupted. It is easy to reset, follow the instructions below that will guide you through the process.

It is recommended to also recheck the Letter Calibration of your letter size. Follow the steps below:-

  1. Go to the top toolbar and select “Displays”
  2. select “Change Patient Display”. You might also have to “Switch Toolbar Location”
  3. Select “Save Layout”.
  4. Recheck Calibration as above “Letter size calibration”
  5. Select Options from the top menubar
  6. Select setup wizard
  7. Go To “Step 2 of 4”
  8. Enter 6000 as your room length, and check the tickbox “using a mirror” if you are using one, leave unchecked if not.
  9. On “Step 3” select the Patients screen from the number displayed on the screen click “next”
  10. Measure in Millimetres the height and width of the large plus sign, (ON THE PATIENT MONITOR)  and enter the values
  11. Click “Finish”
  12. Select a 6/60 letter on the VA screen, (Patient Monitor) and measure the height of the letter
  13. Should be 87.5 mm if not, redo height and width of plus sign
  14. If it is 87.5, go back to step 2 and enter Room Length from the patient to the Monitor, and if using a Mirror, ADD this to the length from the mirror to the patient monitor., and check the Mirror tick box on the room length screen.
  15. Select “finish” your letter size is correctly calibrated

Took laptop home now calibration is out

This commonly occurs if a user takes the laptop home, and perhaps watches a movie, or possibly plays a game on the computer, these programs often reset the resolution or aspect ratio.

Please go to Letter Calibration and reset if necessary

New laptop – need to move Spectrum

when purchasing or upgrading a laptop, most often the calibration and the resolution changes, and this can corrupt the settings of your program . It is advisable to go through the steps below to be sure that your settings are correct to maintain the consistency, and accuracy of your program:-

  1. Contact the Call Centre 041 006 1000 (8:00 till 17:00 weekdays)
  2. They will issue a Spectrum download link and Activation Key
  3. Go through the Letter Calibration process

Remote not working

If the Remote stops working one is able to use the program completely on your laptop keyboard and touch pad, or with a mouse. Some of the features of the Spectrum Program will only work with a mouse.

Please follow these steps if your remote stops working:-

  1. Press any button on the remote and see if a light comes on, on the receiver
  2. If yes, unplug the receiver and insert in a different USB port
  3. Close Spectrum and reopen.
  4. If No, check plugs properly inserted and change the remote batteries
  5. If light on the receiver still does not come on, contact “Accounts”, and they will issue you with a new remote

Laptop Stolen, Please Help

If your laptop is stolen or damaged and it is not able to be repaired, you will need to reinstall Spectrum, and your computer will need to be re-calibrated. Please follow the steps below to ensure optimal performance and accuracy of your spectrum program:-

  1. Contact the Call Centre 041 006 1000 (8:00 till 17:00 weekdays)
  2. They will issue a Spectrum download link and Activation Key
  3. Go through the Letter Calibration process

Our technical team might have to Team view into your computer to assess the situation and rectify the problem. Please follow the steps listed below:-

  1. Ensure you have the latest version of “Teamviewer”
  2. Contact the Call Centre 041 006 1000 (8:00 till 17:00 weekdays)
  3. Arrange a convenient time for you to setup teamviewer session

Licence Expired

You might get a notice every day, starting 7 days before your licence is due to expire, however there are situations where you might be on leave, and would not have seen the notifications when opening the spectrum program. If your licence has expired please follow the steps below:-

  1. Contact the Call Centre 041 006 1000 (8:00 till 17:00 weekdays)
  2. Select Option 1

Spectrum Asking for a Key

It might occur that for various reasons your Spectrum Program might ask for an activation key. If this occurs, please follow the steps below:-

  1. Contact the Call Centre 041 006 1000 (8:00 till 17:00 weekdays)
  2. Select Option 1